The Iodine stabilized Helium-Neon laser system is an easy to use calibration system for all high-end wavemeters. Because it's reference frequencies base on the transition of atomic iodine isotop vapour, it is ideal for calibration of our wavelength meters of the Ultimate accuarcy series!
 

Main Benefits and Properties:

  • Fully compliant with the International Committee of Weights and Measures CIPM 97 “Mise en pratique” recommendations for the realization of the meter
  • Frequency repeatability: 2.5 x 10-11
  • Frequency stability: 2.5 x 10-12
  • Compact construction

Technical Information

Laser Head

Nominal output wavelength 633 nm
Frequency stability (10s averaging time) <2.5 x 10-12
Repeatability 2.5 x 10-11
Method of stabilization Third harmonic method
Accessible 127I2 hyperfine components d,e,f,g,h,i,j,k,l,m,n of the 11-5 R(127) absorption
Output power 60 μW (up to 75 µW on request)
Polarization Linear, vertical
Continuous frequency lock over 24 hours
for ambient temperature 20 °C ±1°C
Yes
Tuning Automatic & Manual
Dimensions 460x180x155 mm

Power Supply

AC line voltage 220-240 V / 50 Hz (others on request)
Dimensions 350x250x110 mm
The I2-HeNe can be used as calibration source for WS8-2
WS8
WS8 UV-I
WS8 IR-I
WS7-30
WS7-30 UV-I
WS7-30 IR-I
WS7 IR-I
Older models: WSU2
WSU10
WSU10 UV-I
WSU10 IR-I
WSU30
WSU30 UV-I
WSU30 IR-I
 

Options

Optical Isolator (i) The system can be equipped with an optical faraday isolator to minimize optical feedback and increase the stability of the laser system.

Applications, Notes and Downloads

Applications

HighFinesse Calibration Source

Downloads

  • Calibrations Sources: SLR and others Download

Notes

For further technical information, application examples, diagrams and for customization please contact: info@highfinesse.de

 

Create PDF

Send request