Wavelength meters
WS5
WS6
WS7
WSU
Options
Spectrum Analyzers
LSA
HDSA
Laser Systems
TS:precise
TS:fast
SLR
Current sources
Power classes
Response time
Noise Spectrum
Laser Drivers
InGaAs Camera
|
Laser Spectrum Analyzer - LSA
Absolute accuracy: 6000 MHz, resulution: 1000 MHz
Sensitivity: 5 nJ @ 633 nm
Low stage spectral resolution: (λ/ΔλFWHM): 500
High stage spectral resolution (λ/ΔλFWHM): 20000 (SM fiber),     10000 (50µm fiber)
max. spectral range: 3 THz
linewidth accuracy : 10%
    min. linewidth Δλ/λ= 2*10-5 (e.g. 10pm @ 633nm)
    max. linewidth: 1.5 THz
Built-in light source for autocalibration
NEW: Mid-IR Laser Spectrum Analyzer: LSA IR-III
Click here for more information!
The Laser Spectrum Analyzer is designed to analyze the multi-line or broadband spectrum of light sources like cw and pulsed lasers, gas discharge lamps, super luminescence diodes, semiconductor laser diodes and LEDs.
| Technical Data | |
LSA |
| Measurement range [nm] |
Standard
UV
UV-II
UV-II-Vis
IR-I
IR-II
IR-III (7) |
|
350-1120
248-1100
192-800
192-1190
800-1750
1000-2250
2000-11000 |






 |
|
|
192-370 nm [pm] (1)
370-1100 nm [MHz]
1100-2250 nm [MHz] |
6
6000
12000 |
| Quick coupling accuracy (with MM fiber) [MHz] |
12000 |
| Wavelength deviation sensivity |
|
|
248-370 nm [pm]
370-1100 nm [MHz]
800-1750 nm [MHz] |
5
3000
5000 |
|
|
Standard, UV
IR |
20000(SM),10000(MM)
4000(SM),2000(MM) |
|
|
Accuracy
Min. linewidth [MHz]
Max. linewidth [THz] |
10%
3000 (3)
2 |
Measurement speed [Hz] (5)
(depending on PC hardware and settings) |
|
|
Data aquisition
Wavelength calculation
Linewidth calculation
Pattern display |
400
20
20
15 |
Required input energy [µJ] (or power [µW]) (6) |
|
|
Standard
UV
IR |
0.0001-0.04
0.0001-0.1
0.02-2 |
|
|
FSR (Ghz) |
~5500 |
| Coupling fiber diameter [µm] |
50 µm or SM fiberset |
| Calibration |
Built-in calibration (8) |
| Calibration period |
≤1 month (will be detected and done automatically) |
| Warm-up time |
No warm-up time under constant ambient conditions |
| Dimensions L x W x H [mm] |
325 x 180 x 77 |
| Weight [kg] |
2.8 |
| Interface |
High-speed USB 2.0 connection |
| Power supply |
Power consumption <2.3 W, supply directly via USB cable; IR-III: external power supply included; IR-I and WSU via USB or external power supply possible |
| (1) With multi mode fiber |
  |
(2) Only for standard range of 370-1100nm |
  |
(3) But not better than 5% of the linewidth |
| (4) According to 3σ criteria |
  |
(5) Without auto-calibration usage |
  |
(6) The cw power interpretation in [µW] compares to an exposure of 1s (generally the energy needs to be divided by the exposure time to obtain the required power) |
| (7) For further information on IR-III devices see MID-IR section |
  |
(8) not IR-III, external calibration source needed, e.g. SLR-1532 |
Available Options
» Linewidth option - L
Included in Delivery (standard version)
Optical unit LSA
USB connection cable (optical unit - computer)
1x 50µm Multimode fiber 5 m
2 singlemode fibers on choice (400nm, 460nm, 630nm, 780nm, 980nm avaible)
2x Collimators for coupling light into fibers
CD-ROM with userfriendly software for operation and measurement
Operation manual
Certificate
Links
» Demo Software
» LSA Product Sheet 2012 [PDF, 123 kB]
» LSA IR-III Product Sheet 2012 [PDF, 128 kB]
back to top
|