Home

Products

Distributors

Service

Download
FAQs
eMail

Contact


FAQs - Frequently Asked Questions

What are the axes on the interferometer graphs?
What does "spectral resolution" mean?
What kind of pulsed measurement is possible and what is the accuracy?
What's the difference between good and bad illumination?
Ltr-files - Why?
How to generate an ltr-file?
The Multichannel-Switcher shows the same picture on all channels

What about the temperature stability of the wlm?

Before measuring the wavelength meter needs to be in thermal equillibrium with its environment. Usually it's enough when it's kept in the lab for 1-2 hours before the measurement. As the temperature of the interferometer is measured and considered in the calculation, slow temperature changes doesn't influence the accuracy. But what does "slow" mean?

In the diagrams above you can see the temperature stability of our wlms. The wlm was kept in a climate chamber for several hours while changing the temperature drastically. As long as the change is below approx. 2°C/h, the temperature drift is compensated quite well.

What is the TTL-synchronization option for?
Why is the (re-)calibration of wlm that important?
How to find the best parameters for the PID-controller?

Last update: 12.01.2009 © HighFinesse GmbH 2010